AIMS/BACKGROUND—Recently a commercially available scanning laser Doppler flowmeter has been produced, which provides two dimensional maps of the retinal perfusion. The aim of the present study was to investigate the reproducibility and the sensitivity of these measurements.
METHODS—16 healthy subjects were randomised to inhale different gas mixtures of oxygen and nitrogen in a double blind crossover study. The following gas mixtures of oxygen and nitrogen were administered: 100% oxygen + 0% nitrogen, 80% oxygen + 20% nitrogen, 60% oxygen + 40% nitrogen, 40% oxygen + 60% nitrogen, 30% oxygen + 70% nitrogen, 20% oxygen + 80% nitrogen, 15% oxygen + 85% nitrogen, and 10% oxygen + 90% nitrogen. Retinal haemodynamic variables and systemic haemodynamics were measured during all inhalation periods. Recordings under resting conditions were performed three times to calculate intraclass coefficients.
RESULTS—In two subjects we did not obtain technically adequate results. A dose dependent change in retinal blood flow during graded oxygen breathing was observed (p < 0.001). During 100% oxygen breathing blood flow decrease was between 29% and 33%, whereas blood flow increase was between 28% and 33% during inhalation of 10% oxygen + 90% nitrogen.
CONCLUSIONS—Scanning laser Doppler flowmetry has an acceptable reproducibility and is appropriate for description of the effect of graded changes in PO2 on retinal haemodynamics. The main problems with the system are the large zero offset, the fixation during retinal scanning, and the neglect of blood flow changes during the cardiac cycle.

机译:目的/背景-最近,已经生产出了一种可商购的扫描激光多普勒流量计,该流量计提供了视网膜灌注的二维图。本研究的目的是研究这些测量的可重复性和敏感性。
方法—在双盲交叉研究中,将16名健康受试者随机吸入不同的氧气和氮气混合气体。施用了以下氧气和氮气的气体混合物:100%氧气+ 0%氮气,80%氧气+ 20%氮气,60%氧气 + 40%氮气,40%氧气+ 60%氮气, 30%氧气+ 70%氮气,20%氧气+ 80%氮气,15%氧气+ 85%氮气和10%氧气+ 90%氮气。在所有吸入期均测量视网膜血流动力学变量和全身血流动力学。在静止条件下进行了3次记录,以计算组内系数。
结果-在两个受试者中,我们没有获得足够的技术结果。在分级氧气呼吸期间观察到视网膜血流的剂量依赖性变化(p <0.001)。在进行100%氧气呼吸时,在吸入10%氧气+ 90%氮气的过程中,血流量减少了29%至33%。而结论-扫描多普勒血流仪具有可接受的重现性,适合描述PO2分级变化对视网膜血流动力学的影响。该系统的主要问题是较大的零偏移,视网膜扫描期间的固定以及在心动周期期间血流变化的忽略。