Scalp Blood Flow, Measured by Laser Doppler Flowmetry, and Transcutaneous Po2 and Pco2 in the Lamb

机译:通过激光多普勒血流仪测量的头皮血流量以及羔羊中的经皮Po2和Pco2

获取原文

摘要

To describe the relation between scalp blood flow and transcutaneousPO2 and PCO2, these three signals were recorded simultaneously in eight experiments on four healthy lambs in the first weeks after birth. Scalp blood flow was recorded by a laser Doppler flow sensor that was incorporated in the transcutaneous PO2 electrode. Scalp blood flow was varied by applying a circular pressure on the scalp surrounding the sensors. A decrease in laser Doppler blood flow was associated with a decrease in transcutaneous PO2 and an increase in transcutaneous PCO2, according to a hyperbolic curve. The steep parts of the hyperbolic curves were observed at low scalp blood flow, whereas the flat parts were found at high flow values. Even during heat-induced hyperemia only a small part of these hyperbolic curves showed a reasonable independence of the transcutaneous blood gas values of changes in blood flow. In view of the previously observed decrease in intrapartum scalp blood flow, the present data render the transcutaneous measurement of blood gas values doubtful for fetal monitoring during labor.
机译:为了描述头皮血流量与经皮PO2和PCO2之间的关系,在出生后的头几周内,在四只健康羔羊的八个实验中同时记录了这三个信号。头皮血流量是通过并入经皮PO2电极的激光多普勒流量传感器记录的。通过在传感器周围的头皮上施加圆形压力来改变头皮血流量。根据双曲线,激光多普勒血流的减少与经皮PO2的减少和经皮PCO2的增加有关。在低头皮血流下观察到双曲线的陡峭部分,而在高血流值下观察到平坦的部分。即使在热诱发的充血期间,这些双曲线中的仅一小部分显示出经血流量变化的经皮血气值的合理独立性。鉴于先前观察到的产时头皮内血流减少,本数据使得经皮测量血气值对于分娩过程中的胎儿监测值得怀疑。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号