首页> 美国卫生研究院文献>Sensors (Basel Switzerland) >Excimer Laser Surgery: Biometrical Iris Eye Recognition with Cyclorotational Control Eye Tracker System
【2h】

Excimer Laser Surgery: Biometrical Iris Eye Recognition with Cyclorotational Control Eye Tracker System

机译:准分子激光外科手术:眼动虹膜识别和眼动仪

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A prospective comparative study assessing the importance of the intra-operative dynamic rotational tracking—especially in the treatment of astigmatisms in corneal refractive Excimer laser correction—concerning clinical outcomes is presented. The cyclotorsion from upright to supine position was measured using iris image comparison. The Group 1 of patients was additionally treated with cyclorotational control and Group 2 only with X-Y control. Significant differences were observed between the groups regarding the mean postoperative cylinder refraction (p < 0.05). The mean cyclotorsion can be calculated to 3.75° with a standard deviation of 3.1°. The total range of torsion was from −14.9° to +12.6°. Re-treatment rate was 2.2% in Group 1 and 8.2% in Group 2, which is highly significant (p < 0.01). The investigation confirms that the dynamic rotational tracking system used for LASIK results in highly predictable refraction quality with significantly less postoperative re-treatments.
机译:提出了一项前瞻性比较研究,该评估评估了术中动态旋转跟踪(尤其是在角膜屈光准分子激光矫正术中散光的治疗中)与临床结果相关的重要性。使用虹膜图像比较测量从直立位置到仰卧位置的旋扭。第一组的患者另外接受了环转控制,第二组仅接受了X-Y控制。两组之间的平均术后屈光度有显着差异(p <0.05)。平均旋扭可以计算为3.75°,标准偏差为3.1°。扭转的总范围是从-14.9°到+ 12.6°。第1组的再治疗率为2.2%,第2组的为8.2%,这是非常显着的(p <0.01)。研究证实,用于LASIK的动态旋转跟踪系统可产生高度可预测的屈光质量,而术后的再治疗则明显减少。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号