Separation of the Tip of a Coblation Wand within the Knee Joint: A Complication of Arthroscopic Adhesiolysis

机译:膝关节内消融棒尖端的分离:关节镜黏附溶解的并发症

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。
获取外文期刊封面目录资料

摘要

Coblation devices are now widely used in arthroscopic surgery and they show a very low incidence of intraoperative complications. We experienced a case where the tip of the wand separated and migrated into the posterior knee compartment in an arthrofibrotic knee. The free wand tip was identified and then extricated from the popliteal hiatus of the knee with using C-arm fluoroscopic control. To the best of our knowledge, this is the first report of its kind involving coblation wands. We describe this complication to show that the use of coblation devices can lead to unexpected problems and it is imperative to inspect all instruments before and after each surgical use.
机译:现在,消融设备广泛应用于关节镜手术中,并且它们的术中并发症发生率非常低。我们经历了一种情况,其中棒的尖端分离并迁移到关节炎性纤维变性膝关节的后膝区。识别出游离的魔杖尖端,然后使用C型臂荧光镜控制从膝盖的hi裂中解脱出来。据我们所知,这是涉及消融魔杖的同类报道中的第一篇。我们对这种并发症进行了描述,以表明使用消融器械会导致意想不到的问题,并且必须在每次手术之前和之后检查所有器械。