Objective:To find a more accurate enzyme standard instrument test method with good repeatability.Methods:The accuracy and repeatability of enzyme standard instrument dual wavelength and single wavelength were compared.Results:The single wavelength of CV was 3.45%.The dual wavelength of CV was 1.86%.The difference was very big.The CV value of the single wavelength was much larger than that of the dual wavelength,but the sensitivity of the two were very same.Conclusion:When using enzyme standard instrument than color,it should be detected using dual wavelength,and then make the accuracy of test analysis more accurately,and reduce the error occurred in the experiment.%目的:寻找一种更加准确和重复性较好的酶标仪检测的方法。方法:对酶标仪双、单波长进行准确性和重复性进行比较。结果:CV的单波长3.45%,CV的双波长1.86%,两者的差距十分大,单波长的CV数值是远远大于双波长,但是两者的灵敏度却是十分相同的。结论:在采用酶标仪比色时,应该使用双波长进行检测,进而使试验分析的准确度更确切,降低实验中出现的误差。