首页> 外文学位 >A comparison of measurement techniques for quality assurance of RapidArc treatment plans.
【24h】

A comparison of measurement techniques for quality assurance of RapidArc treatment plans.

机译:RapidArc处理计划质量保证的测量技术比较。

获取原文
获取原文并翻译 | 示例

摘要

A new form of intensity modulated radiation therapy (IMRT) using the Varian RapidArcRTM treatment system has the potential to improve cancer treatments by delivering comparable dose distributions as TomoTherapy RTM1 at a rate that is 5 to 15 times faster 2. The goal of this thesis was to compare the sensitivity of an electronic portal imaging device (EPID), film, and the ArcCheck(TM) device as tools for evaluation of IMRT treatments. To accomplish this, we introduced systematic errors in MLC leaf position and perform theoretical and experimental evaluations. The comparisons were made by analyzing changes in the gamma function as systematic MLC gap errors were introduced into the patient treatment plan. To study this, dynamic prostate Rapidarc(TM) plans were developed in Eclipse(TM) using the RANDO anthropomorphic phantom.;After the plans were developed, the files were exported in DICOM-RT (Digital Imaging and Communications in Medicine --- Radiation Therapy) format, and a program was written to modify the MLC leaf position. The files were imported back into Eclipse and recalculated using simulated film, EPID, and ArcCheck phantoms. Before measuring the plans with the devices, each instrument was calibrated and repeatability tests were performed to determine the variation in a single plan. After the delivery of the same plan eight times, film was found to have the largest variation in average gamma of 0.31 +/- 0.13. When the setup was not moved in between measurements, the ArcCheck and EPID had significantly smaller variations in average gamma of 0.10 +/- 0.04 and 0.07 +/- 0.03 respectively. When the setup was moved and realigned in between measurements, the average gamma variation was found to be 0.22 +/- 0.10 and 0.11 +/- 0.05 for the ArcCheck and EPID. As a result of the higher variation, it was decided that testing should be performed without moving the detector during the course of the measurements. Film, unfortunately, had to be changed in between tests, which resulted in a high level of variation. After calibration, the original five plans and their modified MLC gap variants were tested.;A comparison of the Eclipse plans with changes in leaf position to the original plan with 0 mm change in gap shows that the film should have the least sensitivity to changes in average gamma. When looking at the measured results with changes in MLC gap compared to the measured plan with 0 mm change in leaf position, we see that film incorrectly appears to have the highest sensitivity. The cause of this was due to the large variation in average gamma discovered during the repeatability tests. By comparing the repeatability value of 0.31 +/- 0.13 to the sensitivity curve generated by Eclipse, it is estimated that film cannot accurately measure changes in systematic gap less than 2 mm. A similar value was found by Yan et al. 15 when using the 2 % / 2 mm gamma <= 1 pass rate.;The EPID and ArcCheck displayed significantly less variation in average gamma during static tests (0.07 +/- 0.03 and 0.10 +/- 0.04 respectively). Using the Eclipse calculated graph showing changes in average gamma compared to changes in MLC gap, we find that the EPID and ArcCheck should be able to measure systematic MLC errors of greater than 0.3mm. Previous analysis on MLC motion indicates that most positional errors are below 0.5 mm16 , which makes it difficult to conclude if the repeatability of the device is limited by the detector, or by actual errors in MLC leaf position on the treatment machine. For systematic gap width changes in the range of 0 to 1 mm, the EPID and ArcCheck show similar sensitivity.;This thesis has focused on how the average gamma value changes when comparing systematic changes in MLC gap to their original unchanged plan. Although the results are not conclusive, they do indicate that further testing may be warranted. Future testing should involve other treatment sites with steeper dose gradients such as head and neck plans. To ensure clinical relevancy, measurements should also focus on smaller systematic errors based on actual expected MLC gap deviations. If further testing does confirm that sensitivity does change with the geometric shape of the measurement device, it could potentially have an influence on future instrument designs. (Abstract shortened by UMI.).
机译:使用Varian RapidArcRTM治疗系统的一种新形式的调强放射疗法(IMRT)具有以与TomoTherapy RTM1相当的剂量分布,并且以快5至15倍的速度提供了与之相当的剂量分布。2。本论文的目标是比较电子门禁成像设备(EPID),胶片和ArcCheck™设备作为评估IMRT治疗的工具的灵敏度。为此,我们在MLC叶片位置中引入了系统误差,并进行了理论和实验评估。通过将系统性MLC间隙错误引入患者治疗计划中,通过分析伽马功能的变化进行比较。为了研究这一点,使用RANDO拟人模型在Eclipse™中开发了动态前列腺Rapidarc™计划。在计划制定之后,将文件导出到DICOM-RT(医学数字成像和通信---辐射)中。疗法)格式,并编写了修改MLC叶位置的程序。这些文件被导入回Eclipse,并使用模拟胶片,EPID和ArcCheck幻象重新计算。在使用设备测量计划之前,先对每台仪器进行校准,然后进行重复性测试以确定单个计划中的变化。在交付了相同的计划八次后,发现胶片的平均伽玛值变化最大,为0.31 +/- 0.13。当两次测量之间没有移动设置时,ArcCheck和EPID的平均伽玛值分别显着较小,分别为0.10 +/- 0.04和0.07 +/- 0.03。在两次测量之间移动设置并重新对齐时,发现ArcCheck和EPID的平均伽玛变化为0.22 +/- 0.10和0.11 +/- 0.05。由于变化较大,因此决定在测量过程中应在不移动检测器的情况下进行测试。不幸的是,在两次测试之间必须更换胶片,这导致了很大的差异。校准后,测试了原始的五个平面图及其修改后的MLC间隙变型。将叶片位置变化的Eclipse平面图与间隙变化为0 mm的原始平面图进行比较,结果表明该膜对变化的敏感性最低。平均伽玛。当观察具有MLC间隙变化的测量结果与叶片位置变化为0 mm的测量计划相比时,我们发现胶片错误地具有最高的灵敏度。其原因是由于在重复性测试中发现的平均伽马值存在较大差异。通过将0.31 +/- 0.13的可重复性值与Eclipse生成的灵敏度曲线进行比较,可以估计胶片无法准确测量小于2 mm的系统间隙的变化。 Yan等人发现了类似的值。当使用2%/ 2 mm伽玛<= 1通过率时为15; EPID和ArcCheck在静态测试期间显示的平均伽玛变化明显较小(分别为0.07 +/- 0.03和0.10 +/- 0.04)。使用Eclipse计算得出的显示相对于MLC间隙变化的平均伽玛变化的图表,我们发现EPID和ArcCheck应该能够测量大于0.3mm的系统性MLC误差。先前对MLC运动的分析表明,大多数位置误差都在0.5 mm16以下,这使得很难得出该设备的可重复性是否受到检测器或治疗机上MLC叶片位置实际误差的限制。对于系统间隙宽度在0到1 mm范围内的变化,EPID和ArcCheck表现出相似的灵敏度。本论文着重研究了将MLC间隙系统变化与其原始不变计划进行比较时平均伽马值如何变化。尽管结果不确定,但确实表明可能需要进一步测试。将来的测试应包括剂量梯度较陡的其他治疗部位,例如头颈计划。为确保临床相关性,测量还应基于实际预期的MLC间隙偏差集中于较小的系统误差。如果进一步的测试确实确认灵敏度确实会随着测量设备的几何形状而变化,则可能会对将来的仪器设计产生影响。 (摘要由UMI缩短。)。

著录项

  • 作者

    Konieczny, Jeff.;

  • 作者单位

    Laurentian University (Canada).;

  • 授予单位 Laurentian University (Canada).;
  • 学科 Physics Radiation.;Biophysics Medical.
  • 学位 M.Sc.
  • 年度 2013
  • 页码 119 p.
  • 总页数 119
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号