首页> 外文期刊>Medical Physics >A study on investigating the delivery parameter error effect on the variation of patient quality assurance during RapidArc treatment
【24h】

A study on investigating the delivery parameter error effect on the variation of patient quality assurance during RapidArc treatment

机译:研究RapidArc治疗过程中传递参数误差对患者质量保证变化的影响

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: The purpose of this study is to evaluate delivery parameter errors (DPEs) and their impact on clinical dose variation with the Varian RapidArc technique.Methods: The dynalog files of 16 head-and-neck patients were retrospectively analyzed to characterize three RapidArc DPEs: dose MU, gantry angle, and MLC gap errors. A total of 64 reconstructed plans were created by creating four variants of each of the original 16 plans (three with the DPEs applied individually and one with the three DPEs combined). These reconstructed plans were compared to the original plans to evaluate the impact of the DPEs on the clinical dose distribution.Results: The mean dose MU, gantry angle, and MLC gap error for all patients were 0.00 ± 0.00 MU, -0.36 ± 0.03°, and 0.00 ± 0.01 mm, respectively. The DPEs had no obvious dosimetric impact on any of the studied dosimetric endpoints except the parotid dose. The gantry angle error, MLC gap error, and combined DPEs changed the parotid Dmean (mean dose) and parotid V30 (volume receiving at least 30 Gy) by l%-2%.Conclusions: It is feasible to use dose distributions reconstructed from dynalog file data as a quality assurance tool. The dose MU, gantry angle, and MLC errors have only minor effects on the accuracy of the delivered dose.
机译:目的:本研究的目的是使用Varian RapidArc技术评估传递参数错误(DPE)及其对临床剂量变化的影响。方法:回顾性分析了16例头颈部患者的动态记录,以表征三种RapidArc DPE。 :剂量MU,龙门角度和MLC间隙误差。通过对原始16个计划中的每个计划创建四个变体,总共创建了64个重构计划(三个分别使用DPE,一个与三个DPE组合)。将这些重建的计划与原始计划进行比较,以评估DPE对临床剂量分布的影响。结果:所有患者的平均剂量MU,机架角度和MLC间隙误差为0.00±0.00 MU,-0.36±0.03°和0.00±0.01 mm。除腮腺剂量外,DPE对任何已研究的剂量学终点均无明显的剂量学影响。台架角度误差,MLC间隙误差和组合的DPE使腮腺Dmean(平均剂量)和腮腺V30(接受至少30 Gy的体积)改变了1%-2%。文件数据作为质量保证工具。剂量MU,龙门架角度和MLC误差对输送剂量的准确性影响很小。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号