This paper deals with system level design considerations for mm-size implantable electronic devices with wireless connectivity. In particular, it focuses on neural sensors as one application requiring such miniature interfaces. Common to all these implants is the need for power supply and a wireless interface. Wireless power transfer via electromagnetic fields is identified as a promising option for powering such devices. Design methodologies, system level trade-offs, as well as limitations of power supply systems based on electromagnetic coupling are discussed in detail. Further, various wireless data communication architectures are evaluated for their feasibility in the application. Reflective impulse radios are proposed as an alternative scheme for enabling highly scalable data transmission at <1pJ/bit. Finally, design considerations for the corresponding reader system are addressed.
机译:本文涉及具有无线连接的MM尺寸可植入电子设备的系统级设计考虑因素。特别是,它专注于神经传感器,作为需要这种微型接口的一个应用程序。所有这些植入物的常见是需要电源和无线接口。通过电磁场传输无线电源传输被识别为供电这些设备的有希望的选项。详细讨论了设计方法,系统级权衡,以及基于电磁耦合的电源系统的限制。此外,对应用程序中的可行性评估各种无线数据通信架构。建议反射脉冲无线电作为在<1PJ /位实现高度可扩展的数据传输的替代方案。最后,解决了对应读取器系统的设计考虑因素。