The dual-wavelength direct spectrophotometric method for uric acid determination in effluent dialysate and deep UV LED based optical spectral sensor for non-invasive online hemodialysis monitoring were developed. The time profiles of uric acid concentration in effluent dialysate were recorded for a group of 12 patients and 88 hemodialysis sessions by means of the sensor. The parameters of the double-pool kinetic model for each session were assessed and analyzed. The feasibility of online optical monitoring in the UV region as a useful and convenient tool for the experimental investigation of uric acid kinetics during hemodialysis was demonstrated.
机译:开发了对富集透析液和深紫色LED基光谱传感器进行尿酸的双波长分光光度法,用于非侵入式在线血液透析监测。 通过传感器记录了一组12名患者和88例血液透析次数的流出透析液中尿酸浓度的时间谱。 评估和分析每个会议的双池动力学模型的参数。 证实了紫外地区在线光学监测作为血液透析期间尿酸动力学实验研究的有用和方便工具的可行性。