Comparison of digital mammograms obtained with the cassette-type retrofit mammography flat panel detector installed on an analog system and the conventional full-field digital mammography

机译:用盒式改装乳房X线摄影平板探测器的数字乳房X线照片的比较安装在模拟系统上和传统的全场数字乳房X线摄影

获取原文

摘要

Purpose: To compare the image quality of digital mammograms obtained with a cassette-type retrofit digital mammography (CRM) flat panel detector (FPD) installed on an analog system and the conventional full-field digital mammography (FFDM). Materials and Methods: Digital mammograms were prospectively obtained with a CRM FPD (RoseM 2430C, DRTECH Corp) installed on an analog system (Lorad M-IV, Hologic) in 90 women (median age, 49 years) who had previous mammograms obtained with the conventional FFDM units. Ninety pairs of mammograms were evaluated by two breast radiologists. The overall image quality (including contrast and sharpness) and visibility of normal structure were evaluated using a 5-point scale (1, poor; 5, excellent). If a lesion presents, the visibility of lesions was evaluated using a 4-point scale (1, not visible; 4: high conspicuity). Results: The contrast and sharpness of CRM FPD (mean, 4.1±0.8 and 4.0±0.9, respectively) were not significantly different from those of FFDM (mean, 4.2±0.5 and 4.2±0.5; P<0.05). Of 90 women, the overall image quality was similar between the two images in 39 (43%); FFDM showed better image quality in 33 (37%); CRM FPD showed better image quality in 18 (20%) (P=0.055). There were 54 lesions (44 calcifications, 6 masses, 3 asymmetries, and one mass with calcifications) in 33 women. The difference of lesion visibility between the FFDM (mean 3.3±0.8) and CRM FPD (mean 3.4±0.8) was not statistically significant (P=0.083). Conclusion: The image quality of the mammograms obtained with CRM FPD was comparable with that of FFDM.
机译:目的:比较用盒式获得的数字乳房X线照片的图像质量改型安装一个模拟系统和传统的全数字化乳腺摄影(FFDM)上数字乳腺(CRM)平板检测器(FPD)。材料和方法:数字乳房X线照片进行前瞻性安装了模拟系统(LORAD M-IV,Hologic公司)的90名妇女(平均年龄49岁)谁曾与获得以前乳房X线照片上一个CRM FPD(RoseM 2430C,DRTECH公司)获得传统FFDM单位。九十对乳房X线照片是由两个乳房放射科医师评估。使用5分制图像的整体质量(包括对比度和清晰度)和能见度正常结构进行了评价(1,差; 5,优)。如果病变礼物,病变的可见性,使用4点标度(;:高显着性4 1,不可见)进行评价。结果:CRM FPD的对比度和清晰度(平均,4.1±0.8和4.0±0.9,)没有从那些FFDM的显著不同(平均,4.2±0.5和4.2±0.5; P <0.05)。 90名妇女的,整体的图像质量是39(43%)的两个图像之间的相似; FFDM显示,33(37%)更好的图像质量; CRM FPD显示,18(20%)较好的图像质量(P = 0.055)。有54个病变(44个钙化,6个质量,3个不对称,并用一个钙化质量)在33名妇女。病变的可见性的FFDM(平均3.3±0.8)和CRM FPD之间的差(平均3.4±0.8)无统计学显著(P = 0.083)。结论:与CRM FPD获得的乳房X线照片的图像质量与FFDM的可比性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号