Application of Sinogram-Based Moving Window Subtraction Technique in OSEM Algorithm to Reduce Ring Artifacts for Cone-Beam Micro-CT Image Quality Assessment

机译:基于SCOMAGA的移动窗口减法技术在OSEM算法中的应用,减少锥形光束微型CT图像质量评估的环形伪影

获取原文

摘要

Severe ring artifacts in cone-beam micro-CT degrade image quality and usually cause error recognition during image-based-observer study or computed-aid-diagnosis (CAD). It causes wrong values, such as size of the tumor. The aim of this study was to derive a moving window subtraction method before the iterative reconstruction step to reduce ring artifacts. To access acceptable image quality data, we used a uniform water cylinder phantom and followed the performance measurement guide, ISO-15708 and EN 16016, to evaluate the imaging performance. Results show there are 1.23 times improvement in SNR and 5% improvement in uniformity. Most importantly, we can reduce the error estimate of the object size from 6 pixels to less than 2 pixels. Our proposed method can efficiently reduce the error rate caused by ring artifacts for automatic size estimation applications.
机译:锥形束微型CT的严重环形伪影降低图像质量,并且通常在基于图像观察者的研究或计算辅助诊断(CAD)期间引起错误识别。它会导致错误的值,例如肿瘤的大小。本研究的目的是在迭代重建步骤之前导出移动窗口减法方法以减少环形伪影。为了访问可接受的图像质量数据,我们使用均匀的水缸模镜,然后采用效率测量指南,ISO-15708和EN 16016,以评估成像性能。结果表明,SNR的提高1.23倍,均匀性提高了5%。最重要的是,我们可以将对象大小的误差估计从6个像素减少到小于2个像素。我们所提出的方法可以有效地降低由环形伪影引起的自动尺寸估计应用引起的错误率。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号