A reflection TIE system consisting of a reflecting microscope and a 4f relay system is presented in this paper, with which the transport of intensity equation (TIE) is applied to reconstruct the three-dimensional (3D) profile of opaque micro objects like wafer structures for 3D inspection. As the shape of an object can affect the phases of waves, the 3D information of the object can be easily acquired with the multiple phases at different refocusing planes. By electronically controlled refocusing, multi-focal images can be captured and used in solving TIE to obtain the phase and depth of the object. In order to validate the accuracy and efficiency of the proposed system, the phase and depth values of several samples are calculated, and the experimental results is presented to demonstrate the performance of the system.
机译:本文提出了由反射显微镜和4F继电器系统组成的反射扎带系统,其中施加强度方程(系列)的传输以重建晶片结构的不透明微对象的三维(3D)轮廓3D检查。由于物体的形状可以影响波的相位,因此可以在不同的重新聚焦平面下用多个相位容易地获取物体的3D信息。通过电子控制的重新焦,可以捕获多焦图像并用于求解绑定以获得对象的相位和深度。为了验证所提出的系统的准确性和效率,计算了几个样本的相位和深度值,并提出了实验结果以证明系统的性能。