The main purpose of this research was to measure the resistance of insertion of a dental implant screw designed based on norm, made of the alloy Ti6A14V [1], which was designed for the cancellous bone density (BD) molar region of the human jaw. The implants were embedded in the in a simulated bone polymer (Sawbone?) [2] and were compared with specimens of the region molar of human jaw cadaver [3]. Using a digital torque wrench, the values of the insertion torque (IT) were calculated in the specimens with the implants inserted [4]. The minimum measured value was 25 Ncm or 0.25 Nm and the density of the maximum torque value measured was 110 Ncm or 1.1 in Nm recorded in the cadaver jawbone. The Pull Out Test (POT) was then carried out in order to determine the resistance of the insertion implant at the bone interface. The experimental results obtained for density, IT and resistance to insertion were then compared in order to determine the strength of fixation of the implant to the bone interface.
机译:本研究的主要目的是测量基于规范的牙科植入螺杆插入的阻力,该牙科螺杆设计为由合金Ti6a14V [1]制成,该牙接为人钳口的松质骨密度(Bd)摩尔地区设计。将植入物嵌入在模拟的骨聚合物(Sawboneα)[2]中,并与人颌骨尸体区域臼齿的标本进行比较[3]。使用数字扭矩扳手,在试样中计算插入扭矩(IT)的值,其中植入物插入[4]。最小测量值为25牛顿厘米或0.25 nm和测得的最大转矩值的密度110牛顿厘米或1.1以Nm记录在尸体颚骨。然后进行拉出试验(罐)以确定插入植入物在骨界面处的电阻。然后比较了密度,与插入抗性的实验结果,以确定植入物到骨界面的固定强度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号