Clinical validation of an in-house EPID dosimetry system for IMRT QA at the Prince of Wales Hospital

机译:威尔士王子IMRT QA内部EPID剂量测定系统的临床验证

获取原文

摘要

In this study a simple method using standard flood-field corrected Electronic Portal Imaging Device (EPID) images for routine Intensity Modulated Radiation Therapy (IMRT) Quality Assurance (QA) was investigated. The EPID QA system was designed and tested on a Siemens Oncor Impression linear accelerator with an OptiVue 1000ST EPID panel (Siemens Medical Solutions USA, Inc, USA) and an Elekta Axesse linear accelerator with an iViewGT EPID (Elekta AB, Sweden) for 6 and 10 MV IMRT fields with Step-and-Shoot and dynamic-MLC delivery. Two different planning systems were used for patient IMRT field generation for comparison with the measured EPID fluences. All measured IMRT plans had >95% agreement to the planning fluences (using 3 cGy/3 mm Gamma Criteria) and were comparable to the pass-rates calculated using a 2-D diode array dosimeter.
机译:在本研究中,研究了使用标准泛洪场校正的电子门户成像装置(EPID)图像进行常规强度调制的放射治疗(IMRT)质量保证(QA)的简单方法。 EPID QA系统在西门子ONCOR印象线性加速器上设计和测试,使用OptiVue 1000st EPID面板(Siemens Medical Solutions USA,Inc,USA)以及具有6个和6个和6个eLekta Ab,瑞典的Elekta Axesse线性加速器10 MV IMRT字段,具有阶梯和拍摄和动态MLC交付。用于与测量的ePID流程进行比较的患者IMRT场生成两种不同的规划系统。所有测量的IMRT计划都有> 95%的规划流量协议(使用3 CGY / 3 mm伽玛标准),并且与使用2-D二极管阵列剂量计计算的传递率相当。