Conventional dose-length product method for CT effective dose evaluation in case of multislice scanners with comparatively large X-ray beam collimation leads to underestimation of effective doses due to partly neglecting of scattered in patient body radiation. It is possible to avoid this problem taking into account "tails" of dose distribution on the length exceeds zone ±50 mm relative to the beam centre. Method of DLP evaluation on the basis of dose distribution curves with "tails" estimated on full length (±200 mm or more) is presented.
机译:用于CT有效剂量评估的常规剂量 - 在具有相对大的X射线束准直的多层扫描仪的情况下,导致有效剂量低估,因为部分忽略了患者体辐射的散射。可以避免此问题考虑到相对于光束中心的长度超过区域±50mm的“尾部”的“尾”。提出了基于多长度(±200mm或更大)的“尾”的剂量分布曲线的DLP评估方法。