The contribution of error of sensor for atmospheric pressure /Δpapprox=1 mm Hg/ to expanded uncertainty of measurement of 1 m length using laser interferometer is examined. A prognosis of uncertainty is drawn up. The air refraction ratio is determined using refractometer before and after measurements. The values of this coefficient could be replaced with theoretically determined ones, calculated on the results of measurement of atmospheric pressure using appropriate sensor and the uncertainty sensor's calibration certificate. New uncertainty budget could be drawn up. The results of research demonstrate that contribution to uncertainty of measurement of 1 m length is about 0, 217 μm at atmospheric pressure error 1 mm Hg.
机译:检查传感器误差/Δpappapprox= 1mm Hg /以扩大使用激光干涉仪的测量的不确定度。制定了不确定性的预后。在测量之前和之后使用折射仪测定空气折射率。可以使用适当的传感器和不确定性传感器的校准证书更换在理论上确定的系数的值。可以制定新的不确定性预算。研究结果表明,在大气压误差1mm Hg的情况下对1米长度的测量不确定度的贡献约为0,217μm。