This paper presents a numerical analysis of a novel external support device for coronary artery bypass graft (CABG). The analysis includes the fluid and the structural dynamics of vein grafts with and without an external support device (ESD). Six different vein grafts geometries were modeled under time-dependent physiological conditions. The hemodynamics and the stresses developed at the graft wall were analyzed on patient based geometries 12 months posttransplantation. Results: veins under constriction of ESD were associated with higher wall shear stresses, lower shear gradients, uniform flow and lower structural effective stresses. In conclusion, these findings suggest that constricted graft with ESD improves hemodynamic and structural factors which are correlated to graft failure following CABG procedure.
机译:本文提出了一种用于冠状动脉旁路移植物(CABG)的新型外部支撑装置的数值分析。该分析包括具有和没有外部支撑装置(ESD)的静脉移植物的流体和结构动态。在时间依赖性生理条件下建模六种不同的静脉移植物几何。分析了移植物壁上发育的血流动力学和应力对持续后的患者的几何形状12个月。结果:ESD收缩下的静脉与更高的壁剪切应力,较低的剪切梯度,均匀的流动和较低的结构有效应力相关。总之,这些研究结果表明,具有ESD的收缩移植改善了与CABG程序后与移植失败相关的血流动力学和结构因素。