Ultrasound Assessment of Dental Implant Stability: Finite Element Analysis of Wave Propagation

机译:牙科植入物稳定性的超声评估:波传播有限元分析

获取原文

摘要

Dental implants are widely used in oral and maxillofacial surgery. In clinics, accurate measurements of implants biomechanical stability are of interest since they could be used to improve the surgical strategy by adapting the choice of the healing period to each patient. Experiments using 10 MHz ultrasound device show significant effects of the amount and quality of bone in contact with the implant on the ultrasonic responses measured. In this paper, we present some numerical results for studying the sensitivity of the ultrasonic response of the implant to variations of the quantity and quality of bone tissue in contact with the implant is assessed under realistic conditions. A 3D finite element model which is based on experimental measurements was developed in order to account for the evolution of the bone biomechanical properties at the implant interface. Based on the reproducibility of the measurement, the results indicate that the device should be sensitive to the effects of a healing duration of less than one week. In all cases, the amplitude of the implant response is shown to decrease when the dental implant primary and secondary stability increase, which is consistent with the experimental results. Moreover, the effect of osseointegration phenomena in a layer between bone tissue and the implant is modeled. The numerical results show that the implant ultrasonic response changes significantly when a liquid layer is located at the implant interface compared to the case of an interface fully bounded with bone tissue.
机译:牙种植体广泛用于口腔和颌面外科。在诊所中,精确测量植入物生物力学稳定性非常感兴趣,因为它们可以通过适应每个患者的选择来改善手术策略。使用10MHz超声装置的实验显示出与测量超声响应的植入量接触的骨骼的量和质量的显着效果。在本文中,我们给出一些数值结果为研究在现实条件下被评估植入物在与植入物接触的量和骨组织的质量的变化的超声波响应的灵敏度。开发了一种基于实验测量的3D有限元模型,以便考虑植入界面处的骨生物力学性质的演变。基于测量的再现性,结果表明该装置应对愈合持续时间的效果敏感,愈合持续时间不到一周。在所有情况下,当牙科植入物初级和二次稳定性增加时,植入物反应的幅度被显示为减少,这与实验结果一致。此外,建模了骨组织与植入物之间的骨整合现象的影响。数值结果表明,与用骨组织完全绑定的界面的情况相比,植入物超声波响应在植入界面位于植入界面时变化显着。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号