A radio-photoluminescence (RPL) dosimeter with a copper-ion luminescent center was fabricated to evaluate its response in ionized particle detection. A focused proton microbeam with varying energies up to 3 MeV and heavy ions of 490 MeV osmium (Os) were employed along with X-rays to evaluate its performance in micrometer-scale radiation monitoring. The response to ionized particles was evaluated under focused proton beam irradiation where the peak wavelength differed from that obtained under X-ray irradiation. Two peaks were observed under Os irradiation where the secondary-generated particles and photons have a significant effect on the dosimeter. The results suggest that the fabricated RPL dosimeter with copper luminescence center could be used to estimate the irradiation effect of primary ionized particles separately from the effects of secondary particles, photons, and environmental background radiation.
机译:制备具有铜离子发光中心的无线电光致发光(RPL)剂量计,以评估其在电离颗粒检测中的响应。具有不同能量的聚焦质子微沟,高达3meV和490meV锇(OS)的重离离子以及X射线,以评估其在微米级辐射监测中的性能。在聚焦质子束照射下评价对电离颗粒的响应,其中峰值波长与在X射线照射下获得的峰值波长不同。在OS辐射下观察到两个峰,其中二次产生的颗粒和光子对剂量计具有显着影响。结果表明,具有铜发光中心的制造的RPL剂量计可用于分开估计二次粒子,光子和环境背景辐射的效果的初级电离颗粒的辐照效应。