Exploring Data-Dependent Acquisition Strategies with the Instrument Control Libraries for the Thermo Scientific Instruments

机译:利用仪器控制库为热科学仪器探索数据相关的采集策略

获取原文

摘要

The main advantage of interfacing directly with the on-board computer of the mass spectrometer during data collection is to be able to use the incoming information in real time to direct future decisions. At this point we have attempted several approaches to make additional decisions during data dependent acquisition. Intensity-based dynamic exclusion considers the change in the intensity of the target precursor ion peak in order to perform fragmentation near the peak’s maximum. On-the-fly isotopic envelope extraction from the full scan allows us to select only the monoisotopic ions for fragmentation. Full scan de-isotoping and selection of the monoisotopic species is a prerequisite for efficient permanent exclusion or permanent inclusion strategies because the overall fidelity of these strategies strongly depends not only on the mass accuracy of the instrument but also on the ability to exclude the heavy isotopes from being considered as targets for fragmentation in future runs.
机译:在数据收集期间直接与质谱仪的车载计算机接口的主要优点是能够实时使用来电以指导未来的决策。此时,我们尝试了几种方法在数据相关采集期间进行额外的决策。基于强度的动态排除期考虑了目标前体离子峰值强度的变化,以便在峰值最大值附近进行碎片化。从完全扫描中提取的飞行同位素包络允许我们只选择单位素过离子进行碎片化。全扫描去体外和选择单位缺乏物种是有效的永久排除或永久包容策略的先决条件,因为这些策略的整体保真度不仅取决于仪器的大规模准确性,而且还取决于排除沉重同位素的能力被认为是未来跑步的碎片目标。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号