首页> 外文会议>ASMS Conference on Mass Spectrometry and Allied Topics >Systematic investigation of the failure of one analyte during method transfer and cross check of a 4-in-1 LC-MS/MS method
【24h】

Systematic investigation of the failure of one analyte during method transfer and cross check of a 4-in-1 LC-MS/MS method

机译:系统转移期间一种分析物失效的系统研究,并交叉检查的4 in-1 LC-MS / MS方法

获取原文

摘要

The root cause was identified to be a non-specific binding (NSB) of M1 to glass vial. Because the NSB of M1 in the working solution is minor (-30.2 to -21.2% after 5 consecutive transfers, table 4) and only will negatively impact the signal noise ratio at LLOQ a revalidation of the established method was not required. New cross check experiment was conducted with M1 working solution prepared in polypropylene tube and the results passed with mean difference of 2.7% (28 out of 29 incurred samples selected, the difference for the failed one is -22.7%, table 5). Detailed method description and thorough review are critical during method transition.
机译:根本原因被鉴定为M1的非特异性结合(NSB)至玻璃小瓶。因为工作解决方案中的M1中的NSB是次要的(连续转移后的5-30.2至-21.2%,表4),并且只有将对LLOQ的信号噪声率产生负面影响,因此不需要既定方法的重新验证。用聚丙烯管中制备的M1工作溶液进行新的交叉检查实验,并通过平均差异的结果为2.7%(29个产生的样品中的28个,失败的差异为-22.7%,表5)。详细方法描述和彻底的审查在方法转换期间是至关重要的。