首页> 外文会议>ASMS Conference on Mass Spectrometry and Allied Topics >Evaluating and Improving the Stability of Nanospray for Operation at μL/min Flow Rates with an Integrated Chip-Based Solution
【24h】

Evaluating and Improving the Stability of Nanospray for Operation at μL/min Flow Rates with an Integrated Chip-Based Solution

机译:评价和提高纳米普通稳定性的μl/ min流速与基于集成芯片的溶液的稳定性

获取原文

摘要

Confirmed that the PicoChip column format can be used in nanoflow as well as for microflow applications Demonstrated PicoChip column compatibility at flow rates up to 6 μL/min. Observed 37% decrease in maximum column pressure from 6,594 psi at ambient to 4,172 psi; at 50°C at 4 μL/min. Reduced the mass spectrometer cycle time by 50% from 30 min. per injection to 15 min. per injection at 5 μL/min.
机译:确认Picochip柱形格式可用于纳米卷,以及微射线应用,以6μl/ min的流速显示皮卡坡柱相容。在环境温度下从6,594 psi的最大柱压力观察到37%降低至4,172 psi;在4μl/ min的50℃下。将质谱仪循环时间降低50%从30分钟。每次注射到15分钟。每次注射为5μl/ min。