Application of PENELOPE code to the efficiency calibration of coaxial germanium detectors

机译:Penelope码在同轴锗探测器效率校准中的应用

获取原文

摘要

The code PENELOPE is applied here to the efficiency calibration exercise described in the EUROMET project 428 [Lepy et al., 2001.Appl. Radiat. Isot. 55, 493]. We evaluated the peak efficiencies for a coaxial HPGe detector in the range 60-2000 keV, for point sources located at various distances from the detector and for a cylindrical box containing two different matrixes: a low-density silica and a hydrochloric acid solution. After an optimization of several detector parameters, the results were in good agreement with the experimental values.
机译:代码Penelope在此处应用于Eutomet项目428中描述的效率校准运动[Lepy等,2001.appl。 radiat。 Isot。 55,493]。我们评估了在60-2000keV范围内的同轴HPGE检测器的峰值效率,用于位于距检测器的各个距离的点源,以及包含两个不同基质的圆柱形盒子:低密度二氧化硅和盐酸溶液。在优化几种探测器参数之后,结果与实验值吻合良好。