首页> 外文会议>Annual Scientific Conference of Computers in Cardiology >Examining the Potential of Using Thorax Impedance Measured by Automated External Defibrillators for Quantification of Circulation
【24h】

Examining the Potential of Using Thorax Impedance Measured by Automated External Defibrillators for Quantification of Circulation

机译:检查使用自动外离外除颤器测量的胸部阻抗进行循环量化的潜力

获取原文

摘要

It has been suggested to acquire circulatory information from patients undergoing resuscitation from cardiac arrest by analyzing their thoracic electrical impedance using modified automated external defibrillators (AEDs). To investigate the potential of this idea, we studied the correlation between two impedance-derived parameters related to circulation, the negative peak of the impedance fluctuation (Z{sub}(peak)) and its first time derivative (dZ{sub}(peak)), and arterial blood pressure measurements in 26 patients undergoing resuscitation and 32 hemodynamically stable patients. The highest correlation coefficient, ρ = 0.4338 was found between the systolic blood pressure and the magnitude of the negative peak of the first time derivative of the impedance. The poor correlation indicates that the impedance-derived parameters are not suitable for quantification of circulation, but can be used to indicate circulation.
机译:有人建议通过使用改进的自动除颤器(AED)分析其胸部电阻,从心脏骤停中接受复活的患者的循环信息。为了调查这个想法的潜力,我们研究了与循环相关的两个阻抗导出参数之间的相关性,阻抗波动(Z {sub}(峰))的负峰值及其第一次衍生(DZ {sub}(峰值) ))和动脉血压测量26例经受复苏和32例血流动力学稳定的患者。在收缩压压力和阻抗的第一次衍生物的负峰的负峰值的幅度之间发现了最高的相关系数,ρ= 0.4338。差的相关性表明阻抗衍生的参数不适用于定量循环,但可用于表示循环。