EFFECT OF HEAD SURFACE ROUGHNESS AND STERILIZATION METHOD ON WEAR OF UHMWPE ACETABULAR CUPS: PRELIMINARY HIP JOINT SIMULATOR RESULTS

机译:头表面粗糙度和灭菌方法对UHMWPE髋臼杯磨损的影响:初步髋关节模拟器结果

获取原文

摘要

More than one million hip joint replacements are performed each year in the world. However, the implants do not last forever due to material limitations, even though the operation is successful. The most common material combination used today is a CoCr head articulating against an UHMWPE (Ultra High Molecular Weight Polyethylene) acetabular cup. Several investigations have shown that the metal head is roughened inside the body and thus accelerating wear of the polymer cup. The sterilization method is also known to have effect on the wear properties. In vitro wear tests are however usually performed with as new implants. This investigation aimed at study the impact of head surface roughness on wear of both sterilized and non-sterilized acetabular cups. A total of nine acetabular cup and head pairs were wear tested in a hip joint simulator for 2Million cycles (Mc) with bovine calf serum as lubricant. Wear was determined by weighing of all cups. The heads were of CoCrMo and the average initial head surface roughness was 15nm (R_a), measured with a white light interference profilometer. The roughening was produced with a SiC paper producing circular multidirectional wear tracks to a surface roughness of about 400nm (R_a). The cups were of UHMWPE and the sterilized cups were 3Mrad gamma-radiated in nitrogen. The surface roughness after wear test was unchanged for the roughened heads, while the initially smooth heads were slightly roughened. Preliminary results show that the rough heads increase the wear of the cups 2-fold. The γ-irradiation affected both wear- and soak rate.
机译:每年在世界上进行超过一百万髋关节替代品。然而,即使操作成功,植入物也不会导致由于材料限制而持续。今天使用的最常见的材料组合是铰接针对UHMWPE(超高分子量聚乙烯)髋臼杯的COCR头。几次研究表明,金属头在体内粗糙化并因此加速聚合物杯的磨损。也已知灭菌方法对磨损性能产生影响。然而,通常用新的植入物进行体外磨损试验。这项调查旨在研究头部表面粗糙度对灭菌和非灭菌髋臼杯磨损的影响。总共九个髋臼杯和头部对在髋关节模拟器中进行测试,用于200万次循环(MC),牛犊血清作为润滑剂。通过称量所有杯子来确定磨损。头部是cocroMo,平均初始头表面粗糙度为15nm(r_a),用白色光干型仪测量。用SiC纸产生粗糙的粗糙化,产生圆形多向磨损轨迹,以约400nm(R_a)的表面粗糙度。杯子是UHMWPE,灭菌的杯子在氮气中为3mradγ辐射。磨损试验后的表面粗糙度对于粗糙的头部不变,而最初光滑的头部略微粗糙。初步结果表明,粗糙的头部增加了2倍的杯子。 γ-辐射影响磨损和浸泡率。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号