Physical factors affecting the detection of calcifications in digital mammography

机译:影响数字乳房X线摄影检测的物理因素

获取原文

摘要

We are conducting a perception study to investigate the effects of detector design, magnification, and calcification size on the detection of calcifications in uniform as well as structured background. Calcium carbonate grains of various sizes were used to simulate calcifications. Three detectors were studied: a 100 urn aSi:H/CsI(Tl) flat-panel system, a 30 mu m CCD based system and an analog screen/film combination. The preliminary results show that the flat-panel system with a 100 urn pixel size outperformed a CCD based system with a 30 mum pixel size and an analog screen-film system. However, in studies with an anthropomorphic phantom as the background, the flat-panel system was outperformed by the other two systems. The presence of tissue structure background substantially reduced the calcification detectability. The use of geometric magnification generally improved the detectability with the exception of smallest and largest calcifications.
机译:我们正在进行感知研究,以研究探测器设计,放大和钙化大小对均匀和结构背景下钙化的影响。各种尺寸的碳酸钙颗粒用于模拟钙化。研究了三种探测器:100 URN ASI:H / CSI(TL)平板系统,基于30μMCCD的系统和模拟屏幕/胶片组合。初步结果表明,具有100个URN像素尺寸的平板系统优于基于CCD的系统,具有30毫米像素尺寸和模拟屏幕膜系统。然而,在用拟人虚线作为背景的研究中,平板系统由其他两个系统表现优于其他两个系统。组织结构背景的存在基本上降低了钙化可检测性。除了最小和最大的钙化之外,使用几何放大率通常会改善可检测性。