首页> 外文会议>Mexican Symposium on Medical Physics >Validation of a Custom-made Software for DQE Assessment in Mammography Digital Detectors
【24h】

Validation of a Custom-made Software for DQE Assessment in Mammography Digital Detectors

机译:在乳房X线摄影数字探测器中验证DQE评估的定制软件

获取原文

摘要

This works presents the validation of a custom-made software, designed and developed in Matlab, intended for routine evaluation of detective quantum efficiency DQE, according to algorithms described in the IEC 62220-1-2 standard. DQE, normalized noise power spectrum NNPS and pre-sampling modulation transfer function MTF were calculated from RAW images from a GE Senographe DS (FineView disabled) and a Siemens Novation system. Calculated MTF is in close agreement with results obtained with alternative codes: MTF_tool (Maidment), ImageJ plug-in (Perez-Ponce) and MIQuaELa (Ayala). Overall agreement better than ≈90% was found in MTF; the largest differences were observed at frequencies close to the Nyquist limit. For the measurement of NNPS and DQE, agreement is similar to that obtained in the MTF. These results suggest that the developed software can be used with confidence for image quality assessment.
机译:该作品介绍了在MATLAB中设计和开发的定制软件的验证,旨在根据IEC 62220-1-2标准中描述的算法进行常规评估侦探量子效率DQE。 DQE,归一化噪声功率谱NNP和预采样调制传递函数MTF从GE SANGORPTE DS(FineView Disabled)和西门子初学系统的原始图像计算。计算的MTF与使用替代代码获得的结果密切一致:MTF_TOOL(举例),imagej插件(Perez-Ponce)和Miquaela(Ayala)。在MTF中发现了总体协议优于MTF;在接近奈奎斯特极限的频率下观察到最大差异。对于NNP和DQE的测量,协议类似于MTF中获得的同样。这些结果表明,开发的软件可用于图像质量评估的信心。