首页> 外文会议>European Conference on Solid-State Transducers >CMOS Biomedical Sensor with In Situ Gold Reference Electrode for Urine Detection Application
【24h】

CMOS Biomedical Sensor with In Situ Gold Reference Electrode for Urine Detection Application

机译:CMOS生物医学传感器,原位金参参考电极用于尿检测应用

获取原文

摘要

In this paper, it presents the fully CMOS biomedical sensor with in situ gold reference electrode in the chip. The traditional Ag/AgCl reference electrode is too large to form a biomedical microsystem. Using the on chip gold reference electrode to replace the Ag/AgCl reference electrode has been studied. The ISFET device with native aluminum oxide is used to detect the hydrogen ions. The shape of gold reference electrode is important to get the stable results. To place the surrounding gold reference electrode nearby the sensing membrane is better than the bar-type gold electrode. The sensitivity and linearity of this pH sensor with in situ surrounding gold reference electrode are 56mV/pH and 0.975, respectively. This work demonstrates the optimum gold electrode could be integrated into CMOS-based biomedical sensors and performs the good results.
机译:在本文中,它介绍了芯片中原位金参参考电极的全CMOS生物医学传感器。传统的Ag / AgCl参考电极太大而不能形成生物医学微系统。使用芯片金参比电极替换Ag / AgCl参考电极已经研究过。具有天然氧化铝的ISFET装置用于检测氢离子。金参考电极的形状对于获得稳定的结果是重要的。将周围的金参比电极放置在附近的传感膜优于条形金电极。该pH传感器与原位周围金参比电极的灵敏度和线性分别为56mV / pH和0.975。这项工作证明了最佳金电极可以集成到基于CMOS的生物医学传感器中并进行良好的结果。