Pulse oximetry (PO) is a measurement method used to determine the percentage of blood that is loaded with oxygen (oxygen saturation). It is an easy-to-sue, safe, convenient, inexpensive, painless and noninvasive measurement method for the assessment of oxygen content in the circulatory system and it also allows to measure the pulse frequency (coincident to the heart rate). It is frequently asses in intensive care, in particular in neonatal one and it provides information on the adequacy of cardio/respiratory function. Motion artifacts can largely influence PO measurement performances. The aim of this paper is to compare the PO measurement data obtained (test population: 5 subjects; test duration 2 hours) on fixed limb to the ones obtained in a free-to-move limb conditions, using 2 PO systems applied on the patient limbs. Despite to the manufacturers claimed accuracy data (>2%), we report differences >3.4% on the saturation data between free and fixed limb.
机译:脉搏血氧饱和度(PO)是一种测量方法,用于确定载有氧气的血液百分比(氧饱和度)。它是一种易于使用,安全,方便,廉价,无痛且无创的测量方法,用于评估循环系统中的氧气含量,还可以测量脉搏频率(与心率一致)。它经常在重症监护室中评估,特别是在新生儿中,它提供了有关心脏/呼吸功能是否适当的信息。运动伪影会在很大程度上影响PO测量性能。本文的目的是使用在患者身上应用的两个PO系统,比较在固定肢体上获得的PO测量数据(测试人群:5名受试者;测试持续时间2小时)与在自由活动肢体条件下获得的测量结果进行比较。四肢。尽管制造商声称精度数据(> 2%),但我们报告的自由和固定肢体之间的饱和度数据差异> 3.4%。