Currently, in Brazil, from all computed radiography (CR) systems to mammography exams, there are 160 Fuji film Computed Radiography Mammography System (FCR-Profect) used with different mammographic equipment Until now, this CR is the unique approved by the Food and Drug Administration to use in mammograms. Considering this accreditation, the image quality from 25 mammographic equipment GE and 9 Lorad-Hologic were evaluated using the Fujifilm FCR Mammography QC-Software methodology. This software requires a dedicate phantom named One Shot Phantom M Plus which permits to evaluate the image quality through a single exposure. The results have shown that there are no significant differences of the quality image acquired with GE and Lorad-Hologic equipment based on the following parameters: missed tissue on chest; system sensitivity; geometric distortion, artifacts, uniformity; dynamic range; spatial resolution (SR); SR on magnification condition; low contrast delectability; linearity; additive lag effects; sensibility and multiplicative lag effects (Ghost). From the dose optimizing process the results of relative contrast to noise ratio (CNR) to 20, 40, 60 and 70mm breast PMMA equivalent were similar but, the average glandular dose (AGD) to Lorad-Hologic equipment was bigger than to GE, as follow: 14% to 20mm breast PMMA equivalent; 35% to 40 and 60mm breast PMMA equivalent and 38% to 70mm breast PMMA equivalent. The use of the Fujifilm FCR Mammography QC-Software promotes agility to obtain the experimental data and to analyze the results based on European guidelines for quality assurance in breast cancer screening and diagnosis, fourth edition.
机译:目前,在巴西,所有计算机放射成像(CR)系统,乳腺放射成像检查,有不同的乳房摄影设备中使用到现在为止160富士胶片计算机放射成像乳腺X线摄影系统(FCR-连续红利率),这CR是由食品和药品批准的唯一政府在乳房X线照片使用。考虑到这一点的认可,从25乳腺设备GE和9 LORAD-Hologic的图像质量使用富士FCR乳腺X线摄影QC-软件方法进行评估。此软件需要命名为单次幻影M Plus公司的一个幻象奉献允许通过单次暴露于评价图像质量。结果表明,存在与基于下列参数GE和LORAD-Hologic的设备中获取的图像质量的不显著差异:对胸部错过组织;系统的灵敏度;几何失真,伪影,均匀性;动态范围;空间分辨率(SR); SR上放大的条件;低对比度delectability;线性度;添加剂的滞后效应;感性和可乘滞后效应(虚像)。从剂量优化过程相对对比度与噪声比(CNR)为20,40,60和70毫米乳房PMMA当量的结果相似,但,平均腺体剂量(AGD),以LORAD-Hologic公司设备比到GE更大,如遵循:14%至20mm乳房PMMA当量; 35%至40与60mm乳房PMMA当量,38%的70毫米乳房PMMA当量。使用富士FCR乳腺X线摄影QC-软件促进敏捷性来获得的实验数据和分析基础上对乳腺癌的筛查和诊断,第四版质量保证欧洲指南的结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号