Effects of Eccentric Screwing on the Initial Stability of the Acetabular Cup in Bone of Different Qualities

机译:偏心拧紧对不同品质骨髋穴初始稳定性的影响

获取原文

摘要

Acetabular cup loosening is one of the major failure models of total hip replacement (THR), which is mostly due to insufficient initial stability of the cup. Previous studies have demonstrated that cup stability is affected by the quality of the host bone and the surgical skill when inserting screws. The purpose of this study was to determine the effects on the initial stability of the acetabular cup of eccentric screwing in bone of different qualities. In this study, hemispherical cups were fixed into bone specimens constructed from artificial foam with three elastic moduli using one to three screws. The effects of two types of screw eccentricity (offset and angular) on the cup stability of the acetabular cup were also evaluated. The experimental results indicate that in the presence of ideal screwing, the cup was stable in bone specimens constructed from foam with the highest elastic modulus. In addition, increasing the number of ideal screws enhanced the cup stability, especially in bone specimens constructed from soft foam. Moreover, the cup stability was most affected by offset eccentric screw(s) in the hard-foam bone specimens and by angular eccentric screw(s) in the soft-foam bone specimens. The reported results indicate that the presence of screw eccentricity affects the initial stability of the acetabular cup. Surgeons should keep this in mind when performing screw insertions in THR. However, care is necessary when translating these results to the intraoperative situation due to the experiments being conducted under laboratory conditions, and hence future studies should attempt to replicate the results reported here in vivo.
机译:髋臼杯松动是总髋关节替换(Thr)的主要故障模型之一,主要是由于杯子的初始稳定性不足。以前的研究表明,在插入螺钉时,杯稳定性受到主体骨骼质量和手术技能的影响。本研究的目的是确定对不同品质骨骨椎间偏心螺纹近偏心螺纹的初始稳定性的影响。在这项研究中,将半球形杯固定在由人造泡沫构成的骨标本中,使用一至三个螺钉用三个弹性模量。还评价了两种螺杆偏心(偏移和角度)对髋臼杯的杯稳定性的影响。实验结果表明,在理想的拧紧的情况下,杯子在由具有最高弹性模量的泡沫构成的骨标本中稳定。此外,增加理想螺钉的数量增强了杯稳定性,特别是在由软泡沫构成的骨标本中。此外,杯稳定性受到硬泡沫骨标本中的偏心螺钉和软泡沫骨标本中的角偏心螺钉的影响最大。据报道的结果表明螺杆偏心的存在影响髋臼杯的初始稳定性。在执行THR中的螺钉插入时,外科医生应该记住这一点。然而,在将这些结果转化为术中情况的情况下,由于在实验室条件下进行的实验,因此未来的研究应该尝试复制在此处报告的结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号