首页> 外文会议>World congress on medical physics and biomedical engineering;International congress of the IUPESM >Effects of Eccentric Screwing on the Initial Stability of the Acetabular Cup in Bone of Different Qualities
【24h】

Effects of Eccentric Screwing on the Initial Stability of the Acetabular Cup in Bone of Different Qualities

机译:偏心旋拧对不同骨质髋臼杯初始稳定性的影响

获取原文

摘要

Acetabular cup loosening is one of the major failure models of total hip replacement (THR), which is mostly due to insufficient initial stability of the cup. Previous studies have demonstrated that cup stability is affected by the quality of the host bone and the surgical skill when inserting screws. The purpose of this study was to determine the effects on the initial stability of the acetabular cup of eccentric screwing in bone of different qualities. In this study, hemispherical cups were fixed into bone specimens constructed from artificial foam with three elastic moduli using one to three screws. The effects of two types of screw eccentricity (offset and angular) on the cup stability of the acetabular cup were also evaluated. The experimental results indicate that in the presence of ideal screwing, the cup was stable in bone specimens constructed from foam with the highest elastic modulus. In addition, increasing the number of ideal screws enhanced the cup stability, especially in bone specimens constructed from soft foam. Moreover, the cup stability was most affected by offset eccentric screw(s) in the hard-foam bone specimens and by angular eccentric screw(s) in the soft-foam bone specimens. The reported results indicate that the presence of screw eccentricity affects the initial stability of the acetabular cup. Surgeons should keep this in mind when performing screw insertions in THR. However, care is necessary when translating these results to the intraoperative situation due to the experiments being conducted under laboratory conditions, and hence future studies should attempt to replicate the results reported here in vivo.
机译:髋臼杯松弛是全髋关节置换(THR)的主要失败模型之一,这主要是由于杯的初始稳定性不足所致。先前的研究表明,杯的稳定性受插入螺钉的宿主骨质量和手术技能的影响。这项研究的目的是确定不同质量的骨对偏心螺钉髋臼杯的初始稳定性的影响。在这项研究中,将半球形的杯子固定在由人造泡沫制成的骨骼标本中,该人造标本使用一到三个螺钉具有三个弹性模量。还评估了两种类型的螺钉偏心距(偏移和角度)对髋臼杯稳定性的影响。实验结果表明,在理想的螺纹连接下,该杯在由具有最高弹性模量的泡沫制成的骨骼样本中是稳定的。此外,增加理想螺钉的数量可以增强杯的稳定性,尤其是在用软泡沫制成的骨标本中。此外,硬泡沫骨标本中的偏心偏心螺钉和软泡沫骨标本中的偏心偏心螺钉对杯的稳定性影响最大。报告的结果表明,螺钉偏心距的存在会影响髋臼杯的初始稳定性。在THR中插入螺钉时,外科医生应牢记这一点。但是,由于这些实验是在实验室条件下进行的,因此在将这些结果转换为术中情况时必须格外小心,因此未来的研究应尝试在体内复制此处报道的结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号