PURE MOMENT TESTING FOR SPINAL BIOMECHANICS APPLICATIONS: FIXED VERSUS 3D FLOATING RING CABLE-DRIVEN TEST DESIGNS

机译:螺旋生物力学应用的纯矩测试:固定式与3D浮动环电缆驱动的测试设计

获取原文

摘要

Results show the potential of the fixed ring cable-driven system to deviate from a pure moment loading state. Differences between applied and intended moments were consistently higher for the fixed versus 3D floating ring set-up with the greatest differences in flexion-extension and lateral bending. The fixed ring design also induced non-trivial shear forces. For the 3D floating ring set-up, applied moments remained fairly consistent with intended moments despite varying specimen rigidity and length. Its main advantage is its ability to allow additional directions of unconstrained motion with the linear sliders and vertical bearings on the loading ring.
机译:结果表明,固定环电缆驱动系统有可能偏离纯力矩加载状态。对于固定与3D浮动环装置,施加力矩与预期力矩之间的差异始终较高,并且挠曲-延伸和横向弯曲的差异最大。固定环的设计还引起了不小的剪切力。对于3D浮动环装置,尽管试样的刚度和长度有所变化,但施加的力矩仍与预期力矩保持相当一致。它的主要优点是能够通过加载环上的线性滑块和垂直轴承允许其他不受约束的运动方向。