PURE MOMENT TESTING FOR SPINAL BIOMECHANICS APPLICATIONS: FIXED VERSUS 3D FLOATING RING CABLE-DRIVEN TEST DESIGNS

机译:纯片生物力学应用的纯片试验:固定与3D浮环电缆驱动的测试设计

获取原文

摘要

Results show the potential of the fixed ring cable-driven system to deviate from a pure moment loading state. Differences between applied and intended moments were consistently higher for the fixed versus 3D floating ring set-up with the greatest differences in flexion-extension and lateral bending. The fixed ring design also induced non-trivial shear forces. For the 3D floating ring set-up, applied moments remained fairly consistent with intended moments despite varying specimen rigidity and length. Its main advantage is its ability to allow additional directions of unconstrained motion with the linear sliders and vertical bearings on the loading ring.
机译:结果显示固定环电缆驱动系统偏离纯时刻的加载状态。对于固定的与3D浮动环设置,所应用和预期时刻之间的差异始终较高,具有屈曲 - 延伸和横向弯曲的最大差异。固定环设计还诱导了非琐碎的剪切力。对于3D浮动环设置,尽管样品刚度和长度变化,所施加的时刻仍然与预期的时刻相当一致。其主要优势是其能够允许在装载环上的线性滑块和垂直轴承允许无关紧要的运动的额外方向。