首页> 外文会议>2010 IEEE Nuclear Science Symposium : Conference Record >Comparison of Axial Performance of Cone-Beam Reconstruction Algorithms for Off-Center Flat-Panel Imaging with a SPECT System
【24h】

Comparison of Axial Performance of Cone-Beam Reconstruction Algorithms for Off-Center Flat-Panel Imaging with a SPECT System

机译:SPECT系统用于偏心平板成像的锥束重建算法的轴向性能比较

获取原文

摘要

In this paper, using Defrise Phantoms, we present our investigations of the axial limitations of an analytical cone-beam reconstruction algorithm (FDK), and the iterative ordered-subset transmission reconstruction (OSTR) iterative algorithm, for an axially extended version of the Philips Brightview XCT cone beam CT (CBCT) geometry. Simulations were preformed for head size and body size Defrise Phantoms of different axial dimensions to investigate limitations on axial extent as a function of these. OSTR yielded overall better axial performance than FDK. It may be possible to reconstruct a larger axial extent for brain studies; however, patients with larger body sizes pose more of a problem.
机译:在本文中,使用碎片幻像,我们介绍了对分析锥形束重建算法(FDK)的轴向局限的调查,以及迭代有序 - 子集传输重建(OSTR)迭代算法,用于飞利浦的轴向扩展版本BrightView XCT锥梁CT(CBCT)几何形状。对于头部尺寸和体尺寸的仿真进行了预先成形的不同轴向尺寸的幽灵,以研究轴向范围的限制。 OSTR比FDK更好地产生更好的轴向性能。可以在脑研究中重建更大的轴向范围;然而,体尺寸较大的患者造成了更多的问题。