首页> 外文会议>Medical Imaging 1997: Physiology and Function from Multidimensional Images >Left ventricle myocardial border detection in three-dimensional intracardiac ultrasound images
【24h】

Left ventricle myocardial border detection in three-dimensional intracardiac ultrasound images

机译:三维心内超声图像中的左心室心肌边界检测

获取原文

摘要

Abstract: We have previously reported an automated approach to detection of endocardial and epicardial borders in individual intracardiac ultrasound (ICUS) images. Here, we report the method's extension to 3D ICUS image data sets. Our method is based on fully automated detection of epicardial and endocardial borders inside a single interactively identified region of interest. BOrder detection is based on an optimal graph-searching approach that utilizes a priori knowledge about left ventricular (LV) anatomy and ultrasound imaging physics. Eight cadaveric pig hearts were used for validation. Two ICUS sequences were obtained from each heart, with a 10 MHz CVIS 10F catheter positioned in the LV across (1) the aortic valve and (2) the mitral valve. Performance of the 3D automated border detection method was assessed by comparing the observer- defined and computer-determined quantitative indices of LV volume and by border positioning errors. The 3D reconstruction of the lV was performed from the sequences of the detected epicardial and endocardial borders using shape- based interpolation and surface rendering. !14
机译:摘要:我们先前已经报道了一种自动方法,用于检测单个心内超声(ICUS)图像中的心内膜和心外膜边界。在这里,我们报告了该方法对3D ICUS图像数据集的扩展。我们的方法基于在单个交互式识别的感兴趣区域内的心外膜和心内膜边界的全自动检测。 BOrder检测基于最佳图搜索方法,该方法利用了有关左心室(LV)解剖和超声成像物理学的先验知识。使用八个尸体猪心脏进行验证。从每个心脏获得两个ICUS序列,将一条10 MHz的CVIS 10F导管放置在LV中穿过(1)主动脉瓣和(2)二尖瓣的位置。通过比较观察者定义和计算机确定的左室容积定量指标以及边界定位误差,评估了3D自动边界检测方法的性能。使用基于形状的插值和表面渲染,从检测到的心外膜和心内膜边界的序列中进行静脉的3D重建。 !14

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号