Effects of Increased Catheter Contact Force on the Ablation Impedance in the Radiofrequency Catheter Ablation of Cardiac Arrhythmias

机译:导管接触力增加对心律失常的射频导管消融射击阻抗的影响

获取原文

摘要

Catheter contact force is well known as one of the factors which has a strong positive correlation with lesion formation. We hypothesized that the catheter contact force substantially affects the ablation impedance variation during radiofrequency (RF) catheter cardiac ablation, due to its direct effect on the electrical current output from the RF generator. However, such a hypothesis needs to be validated through further investigation. Therefore, the purpose of this study was to understand the effect of the catheter contact force related to the contact angle on the ablation impedance in the radiofrequency catheter cardiac ablation process. Thus, we developed a special in vitro experimental system that enables to precisely set the catheter contact angle to the heart tissue's surface. The catheter ablation test was held at five different contact angles to the heart tissue surface. A RF current was delivered at each angle at contact force values of 2, 4, 6, 10, 15, 20, 30, and 40 gf. RF catheter ablation examination was repeated six times each for the five levels of contact angles and eight levels of contact forces. Then, the maximum impedance and average impedance during the ablation process for each condition were recorded. The results revealed that the catheter contact force has a significant correlation with ablation impedance. When the contact force was increased, the different impedance between the maximum impedance and average impedance also increased. The percentage of contact area also has a significant correlation with the ablation impedance. When the percentage of contact area was increased, the different impedance between the maximum and the average impedance also increased. However, the ablation impedance did not have a significant difference with each catheter contact angle. The present study showed a significant impact of the catheter contact force on the ablation impedance in RF catheter ablation systems.
机译:导管接触力是众所周知的一种因素之一,其具有与病变形成强的正相关。我们假设导管接触力基本上影响射频(RF)导管心脏消融期间的消融阻抗变化,因为它直接对来自RF发生器的电流输出的电流。但是,需要通过进一步调查来验证这样的假设。因此,本研究的目的是了解导管接触力与接触角相关的导管接触力对射频导管心脏消融过程中的消融阻抗有关。因此,我们开发了一种特殊的体外实验系统,使得能够精确地将导管接触角与心脏组织的表面精确设置。导管消融测试以五个不同的接触角保持在心脏组织表面。在接触力值为2,4,6,10,15,20,30和40gF的接触力值下以每个角度递送RF电流。 RF导管烧蚀检查每次重复六次,每个接触角和八个水平的接触力。然后,记录了每个条件的消融过程中的最大阻抗和平均阻抗。结果表明,导管接触力与消融阻抗具有显着的相关性。当接触力增加时,最大阻抗和平均阻抗之间的不同阻抗也增加了。接触面积的百分比也与消融阻抗具有显着的相关性。当接触面积的百分比增加时,最大值和平均阻抗之间的不同阻抗也增加。然而,消融阻抗与每个导管接触角没有显着差异。本研究表明,导管接触力对RF导管消融系统中的消融阻抗的显着影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号