首页> 外文会议>IEEE International Conference on Micro Electro Mechanical Systems >REAL-TIME MONITORING OF A SELECTIVE FILTER USING MICROCHIP FLOW CYTOMETRY INTEGRATED POLYMER WAVEGUIDES
【24h】

REAL-TIME MONITORING OF A SELECTIVE FILTER USING MICROCHIP FLOW CYTOMETRY INTEGRATED POLYMER WAVEGUIDES

机译:使用微芯片流式细胞术集成聚合物波导的选择性滤波器的实时监测

获取原文

摘要

Both dielectrophoresis (DEP) and flow cytometry are widely applied in microfluidic systems for the handling and measurements of cells and particles. Here, we present a novel integrated microchip, combining a DEP selective filter with flow cytometers for real-time monitoring of the cell sorting processes. The DEP filter was integrated inside a microfluidic channel to sort cells by positive DEP. The cells were hydrodynamically focused to the center of the channel by using a sheath flow from either side. Two flow cytometer (FC) detection windows were set upstream and downstream, respectively, from the DEP filter to measure the forward scattered light of passing cells. Comparing the cell counting rates measured on two different FCs, the collection efficiency of the DEP filter could be determined. This microchip was used to quantitatively monitor the sorting of viable from nonviable yeast cells.
机译:介电泳(DEP)和流式细胞术均广泛应用于用于处理和测量细胞和颗粒的微流体系统中。在这里,我们提出了一种新的集成微芯片,将DEP选择性过滤器与流式细胞计组合,用于实时监测细胞分选过程。将DEP过滤器整合在微流体通道内以通过正DEP对细胞进行分类。通过使用从任一侧的鞘流,细胞将细胞流体地聚焦到通道的中心。两个流动细胞仪(Fc)检测窗口分别从DEP滤波器设置上游和下游,以测量通过细胞的前向散射光。比较在两个不同FC上测量的小区计数速率,可以确定DEP滤波器的收集效率。该微芯片用于定量监测来自非可行酵母细胞的可行的分选。