Radiation dose in patients submitted to x-ray diagnostic exams at an emergency hospital service

机译:接受急诊医院X射线诊断检查的患者的辐射剂量

获取原文

摘要

Radiographic exams for diagnostic of patients under emergence attendance situations in a hospital aredone with many radiographic parameters as established by the divergence in the biotypes of patients andequipments. The lack of standardization in the exposure parameters (radiation field size, focus-skin distance,tube potential and current-time product) and in the x-ray equipment parameters (total filtration and the targetangle) may cause large variations in the radiation doses of the exposed patients. With the aim to determine theorgan absorbed doses in such patients, two approaches were adopted in this work. The absorbed dose variationsin some organs were estimated with the PCXMC? computational software by simulating exposure conditionsthat are adopted by x-ray operators. Additionally, physical dosimetry with thermoluminescent detectors in theAlderson-Rando? phantom was done. Results showed that there is space for reducing absorbed doses to patientswithout any prejudice to the image quality of the diagnostic exam.
机译:用于在医院急诊时对病人进行诊断的射线照相检查是 根据患者生物型的差异确定了许多射线照相参数 设备。曝光参数(辐射场大小,聚焦皮肤距离, 电子管电势和当前时间乘积)以及X射线设备参数(总过滤量和目标 角度)可能会导致暴露患者的放射剂量发生较大变化。为了确定 此类患者的器官吸收剂量,在这项工作中采用了两种方法。吸收剂量变化 在某些器官中用PCXMC进行了估算?模拟暴露条件的计算软件 X射线操作员采用的方法。此外,在实验室中使用热致发光检测器进行物理剂量测定 Alderson-Rando?幻影完成了。结果表明,存在减少患者吸收剂量的空间 不影响诊断检查的图像质量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号