Assessment of Patient Radiation Dose during Conventional Diagnostic X-ray Examinations in Three Public Hospitals in Northern Jordan using TLDs

机译:用TLDS评估三个公立医院的常规诊断X射线检查中的患者辐射剂量

获取原文
获取原文并翻译 | 示例
           

摘要

This study aims to measure entrance surface doses during routine chest and abdomen x-ray examinations of adult and child patients. Radiation dose measurements were performed using thermoluminescent dosimeters TLD-100s in three major public hospitals in northern Jordan on a total of 100 patients. Wide variations in entrance surface doses were observed within and between hospitals, which might be attributed to significant variations of the selected exposure parameters. For adult patients, the results have shown that the majority of entrance surface dose values from both chest and abdomen examinations were within recommended values of diagnostic reference levels. For child patients, the mean entrance surface dose from chest examinations in three age groups were 0.131 mGy (0-1 y), 0.136 mGy (1-5 y), and 0.191 mGy (5-10 y). These values were considered relatively high compared to the European reference levels and published results in the literature. However, for abdomen examinations, entrance surface dose values were relatively lower than European reference levels. Patient effective doses were estimated using a PCXMC 2.0 Monte Carlo program. The results for both adults and children were found to be relatively lower than the values reported by international publications. Due to the wide variations of entrance surface dose and the higher radiation doses delivered to child patients, this study recommends implementing a quality assurance program in such hospitals to achieve optimization between good image quality and minimum dose according to the as low as reasonably achievable principle. Moreover, the results of this work will provide a useful base for establishing local diagnostic reference levels for chest and abdomen examinations in Jordan.
机译:本研究旨在在成人和儿童患者的常规胸部和腹部X射线检查中测量入射表面剂量。辐射剂量测量是在北约旦的三大公立医院中的热致发光剂量计TLD-100,共有100名患者。在医院内部和医院之间观察到入射表面剂量的宽变化,这可能归因于所选择的曝光参数的显着变化。对于成年患者,结果表明,来自胸部和腹部检查的大多数入射表面剂量值都是在诊断参考水平的推荐值范围内。对于儿童患者,来自三个年龄组的胸部检查的平均入射表面剂量为0.131变化(0-1 y),0.136米(1-5 y)和0.191米(5-10 y)。与欧洲参考水平相比,这些值被认为是相对较高的,并在文献中公布了结果。然而,对于腹部检查,入射表面剂量值比欧洲参考水平相对较低。使用PCXMC 2.0 Monte Carlo程序估算患者有效剂量。发现成年人和儿童的结果比国际出版物报告的价值观相对较低。由于入射表面剂量和较高的辐射剂量传递给儿童患者的较大变化,本研究建议在这种医院实施质量保证计划,以便根据合理可实现的原则,在良好的图像质量和最小剂量之间实现优化。此外,这项工作的结果将为建立胸部和腹部检查的局部诊断参考水平提供有用的基础。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号