ESTIMATION OF PATIENT DOSES FOR COMMON DIAGNOSTIC X-RAY EXAMINATIONS IN LATVIAN HOSPITALS, ANALYSIS OF RADIOGRAPHIC TECHNIQUES AND COMPARISON WITH EUROPEAN GUIDELINES

机译:拉脱维亚医院X线检查诊断患者的剂量估计,放射线照相技术分析以及与欧洲指南的比较

获取原文
获取原文并翻译 | 示例
           

摘要

This article deals with the estimation of doses received by patients undergoing radiological examinations in order to establish dose reference levels (DRLs) in Latvia. Several large hospitals, small hospitals and private practices were selected for patient dose measurements. The measurements were carried out using calibrated thermoluminiscence dosemeters attached to the patient's skin. Exposure parameters and patient's data were recorded. The entrance surface doses (ESDs) to patients undergoing several common X-ray examinations (chest AP/PA, chest LAT, lumbar spine AP/PA, lumbar spine LAT and pelvis) were measured. Data concerning the kV_p settings, used type of films, focus-film distance and the ESD values were analysed and compared with those recommended by the European Community (EC). Among the different hospitals and private practices, discrepancies in the patient doses and techniques used for the examination were found, where the doses exceeded the EC recommended values owing to a very low kV_p and a very low sensitivity of the screen film combinations used.
机译:本文讨论了接受放射检查以建立拉脱维亚剂量参考水平(DRL)的患者所接受剂量的估计。选择了几家大型医院,小型医院和私人诊所进行患者剂量测量。使用附在患者皮肤上的校准热发光剂量计进行测量。记录暴露参数和患者数据。测量了接受几次普通X射线检查(胸部AP / PA,胸部LAT,腰椎AP / PA,腰椎LAT和骨盆)的患者的入口表面剂量(ESDs)。分析了有关kV_p设置,使用过的胶片类型,聚焦胶片距离和ESD值的数据,并与欧洲共同体(EC)推荐的数据进行了比较。在不同的医院和私人诊所中,发现用于检查的患者剂量和技术存在差异,这是由于kV_p极低且所使用的屏幕胶片组合的灵敏度极低,因此剂量超过了EC推荐值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号