首页> 外文期刊>Radiation Protection Dosimetry >PATIENT DOSE, IMAGE QUALITY AND RADIOGRAPHIC TECHNIQUES FOR COMMON X RAY EXAMINATIONS IN TWO GREEK HOSPITALS AND COMPARISON WITH EUROPEAN GUIDELINES
【24h】

PATIENT DOSE, IMAGE QUALITY AND RADIOGRAPHIC TECHNIQUES FOR COMMON X RAY EXAMINATIONS IN TWO GREEK HOSPITALS AND COMPARISON WITH EUROPEAN GUIDELINES

机译:两种希腊医院常用X射线检查的患者剂量,图像质量和放射线照相技术,以及与欧洲指南的比较

获取原文
获取原文并翻译 | 示例

摘要

The radiographic technique factors and the quality of each radiographic image for three common examinations (chest PA, pelvis AP and lumbar spine LAT) were compared with the European criteria and entrance surface dose (ESD) was measured For each radiograph in two Greek hospitals. The measurements were carried out using calibrated LiF thermoluminescence dosemet- Ers. The patients were selected so that their weight was close to 70 ± 10 kg and their height to 170 ± 10 cm. Effective dose Values were calculated using a PC-based Monte Carlo program. All four X ray rooms in the survey achieved mean doses well Below the European reference doses. However the mean doses in the X ray rooms differed by a factor of about 3.
机译:将三种常见检查(胸部PA,骨盆AP和腰椎LAT)的放射线技术因素和每张放射线图像的质量与欧洲标准进行了比较,并在两家希腊医院对每张放射线照片测量了入口表面剂量(ESD)。使用校准的LiF热致发光剂量计进行测量。选择患者时,使其体重接近70±10 kg,身高为170±10 cm。使用基于PC的Monte Carlo程序计算有效剂量值。调查中的所有四个X射线室均达到了远低于欧洲参考剂量的平均剂量。但是,X射线室的平均剂量相差约3倍。