Measurement of the Tear Film and Anterior Chamber by Confocal Microscopy

机译:共聚焦显微镜测量泪膜和前房

获取原文

摘要

We report on the development of a novel, low cost instrument that is capable of accurately measuring small, short and long term changes in the thickness of the cornea and tear film at high speed. The performance of the instrument was tested by measuring the influence of Allergan's OPTIVE? lubricating eye drops on the thickness of the cornea and tear film. Comparative measurements to quantify the performance were taken using Haag-Streit's LenStar. It was found that the newly developed instrument accurately measured a change in thickness of around 9 μm with an accuracy comparable to the LenStar, and with a standard deviation of less than 1 micrometer. Since the new instrument was not configured to resolve the tear film from the cornea, we are not yet able to distinguish the cause of the thickening.
机译:我们报告了一种新型低成本仪器的开发,该仪器能够准确地高速测量角膜和泪膜厚度的微小,短期和长期变化。仪器的性能通过测量Allergan的OPTIVE?的影响进行了测试。在角膜和泪膜的厚度上润滑眼药水。使用Haag-Streit的LenStar进行比较测量以量化性能。结果发现,新开发的仪器可精确测量约9μm的厚度变化,其精度可与LenStar媲美,并且标准偏差小于1微米。由于新仪器尚未配置为可解决角膜中的泪膜,因此我们尚无法分辨出增厚的原因。