首页> 外文会议>Italian conference on sensors and microsystems >Improved SQUID Sensors for Biomagnetic Imaging
【24h】

Improved SQUID Sensors for Biomagnetic Imaging

机译:改进的生物磁性成像鱿鱼传感器

获取原文

摘要

Improved niobium-based SQUID sensors in a magnetometer and a gradiometer configuration are presented. The SQUID magnetometer has an integrated square superconducting sensing coil with an area of 9 mm2, much smaller than a typical SQUID magnetometers keeping a comparable magnetic field sensitivity but a higher spatial resolution. At T — 4.2 K, an intrinsic magnetic field noise spectral density of 5.8 fT/√Hz, has been measured in the flux locked loop configuration. The SQUID planar gradiometer with a long baseline (50 mm) has a pickup antenna consisting of a series of two integrated rectangular coils inductively coupled to the SQUID in a double washer configuration. At T = 4.2 K a magnetic flux noise spectral density of 3 μΦ0/√Hz has been measured. The spectral density of the magnetic field noise referred to one sensing coil, is 3.0 fT/√Hz resulting in a gradient spectral noise of 0.6 fT/cm ? √Hz.
机译:提出了一种磁力计和梯度计配置中的基于铌的鱿鱼传感器。鱿鱼磁仪具有一个集成的方形超导感测线圈,面积为9mm 2,远小于典型的鱿鱼磁力计,其保持可比较的磁场灵敏度,但空间分辨率较高。在T-4.2 k下,在磁通锁定环形配置中测量了5.8英尺/√Hz的固有磁场噪声谱密度。具有长基线(50mm)的鱿鱼平面梯度计具有由一系列两组集成的矩形线圈组成的拾取天线,其在双垫圈配置中电感地连接到鱿鱼。在T = 4.2K处,测量了3μφ0/√Hz的磁通噪声谱密度。磁场噪声的光谱密度称为一个感测线圈,是3.0ft /√Hz,导致梯度光谱噪声为0.6英尺/厘米? √Hz。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号