首页> 外文会议>International Symposium on Biomedical Engineering and Medical Physics >The Differential Oscillometric Method Can Be Used for Recording Beat-to-Beat Arterial Pressure Patterns from Radial Arteries
【24h】

The Differential Oscillometric Method Can Be Used for Recording Beat-to-Beat Arterial Pressure Patterns from Radial Arteries

机译:差分示波法可用于从径向动脉记录搏动动脉压力模式

获取原文

摘要

The aim of this pilot study was to investigate the feasibility of the differential oscillometric technique to record beat-to-beat mean arterial pressure patterns from radial arteries (MAPrad). MAPrad was recorded by an experimental device in 9 healthy subjects during rest, light physical exercise and local cooling. Simultaneously, the finger beat-to-beat mean arterial pressure pattern (MAPfin) and the fingertip skin blood flow by the laser Doppler flowmetry were registered. Results demonstrated that the group-averaged beat-to-beat MAPrad recordings had a high similarity to related MAPfin recordings (r=0.92; range 0.86 to 0.98). However, the dynamic radial-to-finger blood pressure difference was found to vary in some subjects as a result of intensive vasoconstriction. In further development attention should be paid to proper positioning of the local pad-type cuffs to avoid overestimation of the radial pressure.
机译:该试点研究的目的是研究差分示波技术的可行性,以径向动脉(Maprad)记录节拍到搏动的平均动脉压板。 Maprad在休息,轻微的体育锻炼和局部冷却期间,在9个健康受试者中被一个实验装置记录。同时,注册了手指搏动的平均动脉压力图案(MAPFIN)和尖头皮肤血流。结果表明,与相关的Mapfin记录有很高的相似性(r = 0.92;范围0.86至0.98)的高相似性。然而,发现动态径向对手指血压差异在一些受试者中因密集型血管收缩而变化。在进一步的发展中,应注意局部垫型袖口的适当定位,以避免高估径向压力。