Lateral temperature spread of monopolar, bipolar and ultrasonic instruments for robotic assisted laparoscopic surgery

机译:用于机器人辅助腹腔镜手术的单极,双极和超声仪器的横向温度分布

摘要

OBJECTIVE: To assess cautery instruments used in robotic assisted laparoscopic (RAL) surgery in regard to critical heat spread on tissue and cellular level.udMATERIALS AND METHODS: Thermal spread along bovine musculofascial tissues was examined by infrared camera, histology and enzyme assay. Currently used monopolar, bipolar and ultrasonic laparoscopic instruments were investigated at various power settings and application times. The efficacy of using an additional Maryland clamp as a heat sink was evaluated. 45°C was considered the cutoff temperature for possible nerve damage.udRESULTS: Monopolar instruments exhibited a critical spread of 3.5 ± 2.3 mm when applied at 60 W for 1 s. After 2 s, the spread surpassed 20 mm. For adjustable bipolar instruments the spread was 2.2 ± 0.6 mm at 60 W and 1 s and 3.6 ± 1.3 mm at 2 s. PK and LigaSure forceps had spreads of 3.9 ± 0.8 mm and 2.8 ± 0.6 mm respectively, whereas the ultrasonic instrument reached 2.9 ± 0.8 mm. Application of an additional Maryland clamp as a heat sink significantly reduced the thermal spread. Histomorphometric analyses and enzyme assay supported these findings.udCONCLUSION: All coagulation devices used in robotic assisted laparoscopy surgery have distinct thermal spreads depending on power setting and application time Cautery may be of concern due to lateral temperature spread, causing potential damage to sensitive structures including nerves. Our results provide surgeons with a resource for educated decision making when using coagulation devices during robotic procedures.
机译:目的:评估在机器人辅助腹腔镜(RAL)手术中在组织和细胞水平上的临界热扩散所使用的烧灼工具。材料和方法:通过红外热像仪,组织学和酶法检测沿牛肌筋膜组织的热扩散。在各种功率设置和应用时间对当前使用的单极,双极和超声腹腔镜仪器进行了研究。评估了使用额外的马里兰夹作为散热器的功效。 45°C被认为是可能造成神经损伤的临界温度。结果:单极器械在60 W功率下施加1 s时,临界展宽为3.5±2.3 mm。 2 s后,传播超过20毫米。对于可调节的双极仪器,在60 W和1 s时的扩展为2.2±0.6 mm,在2 s时的扩展为3.6±1.3 mm。 PK和LigaSure钳的扩散分别为3.9±0.8 mm和2.8±0.6 mm,而超声仪达到2.9±0.8 mm。附加的马里兰夹钳作为散热器的应用可显着降低热扩散。组织形态计量学分析和酶分析证实了这些发现。 ud结论:机器人辅助腹腔镜手术中使用的所有凝血装置均具有不同的热扩散,具体取决于功率设置和应用时间。侧向温度扩散可能会引起谨慎,可能会对敏感结构造成潜在损害,包括神经。我们的结果为外科医生在机器人手术过程中使用凝血设备时提供了有根据的决策依据。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号