Results of international Dobson spectrophotometer calibrations at Arosa, Switzerland, 1990

机译:1990年瑞士阿罗萨国际多布森分光光度计校准结果

获取原文

摘要

An international comparison of Dobson ozone spectrophotometers, organized and partially funded by the World Meteorological Organization (WMO), was held at the Lichtklimatisches Observatorium (LKO) in Arosa, Switzerland, July-August 1990. Countries participating with a total of 18 Dobson instruments were Belgium, Czechoslovakia, Denmark, Germany, Greece, Hungary, Iceland, Norway, Poland, Portugal, Rumania, Spain, Switzerland, the United Kingdom, the United States, and the United Soviet Socialist Republics. The reference standard instrument for the comparison was U.S.A. Secondary Standard Dobson Spectrophotometer 65 maintained by the NOAA Climate and Monitoring and Diagnostics Laboratory, Boulder, Colorado. The mean difference in ozone obtained with the Dobson instruments relative to Dobson instrument 65, calculated from ADDSGQP observations in the air mass range 1.15-3.2, was minus 1.0 plus or minus 1.2 (1 sigma) percent. The WMO Standard Brewer Spectrometer 39 also participated. In the mean, the Brewer instrument measured 0.6 plus or minus 0.2 (1 sigma) percent more ozone than did Dobson instrument 65. Results are presented, also, of ozone vertical profile measurements made with the Dobson instruments, two Brewer spectrometers, a LIDAR, a balloon ozonesonde flown from Hohenpeissenberg, Germany, and balloon ozonesondes flown from Payerne, Switzerland.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号