The performance of commercially available continuous air monitors used in the workplace to monitor levels of airborne particulates (alpha, beta/photon emitters) was evaluated by personnel at the Pacific Northwest Laboratory. Testing procedures are based on criteria and methods in draft ANSI N42.17B, ''Performance Specifications for Health Physics Instrumentation - Occupational Airborne Radioactivity Monitoring Instrumentation.'' Testing requirements and results are discussed for the following tests: stability; effects of temperature, humidity, and ambient pressure; alarm threshold drift; coefficient of variation; response time; energy dependence; and air flow rate accuracy. Tests results show that certain systems may show large temperature, humidity, or ambient pressure effects; interference effects from radio frequency fields; very slow response times; a large beta-energy dependence; or a saturation problem in high-exposure conditions. Systems performed well when compared to stability, alarm threshold drift, and air flow rate accuracy criteria. 2 refs., 5 figs., 5 tabs. (ERA citation 12:009008)