首页> 外文期刊>Physics in medicine and biology. >T-2, Carr-Purcell T-2 and T-1 rho of fat and water as surrogate markers of trabecular bone structure
【24h】

T-2, Carr-Purcell T-2 and T-1 rho of fat and water as surrogate markers of trabecular bone structure

机译:脂肪和水的T-2,Carr-Purcell T-2和T-1 rho作为小梁骨结构的替代标记

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Magnetic resonance imaging (MRI) techniques have been developed for noninvasive assessment of the structural properties of trabecular bone. These measurements, however, suffer from relatively long acquisition times and low resolution compared to the trabecular size. Spectroscopic measurement of relaxation times could be applied for more detailed and faster assessment of relaxation properties of bone marrow and also provide surrogate information on trabecular structure. In the present study, bovine trabecular bone was investigated with spectroscopic NMR (nuclear magnetic resonance) methods to determine the relationship between structural parameters as measured with micro-CT and T-2, Carr-Purcell T-2 and T-1 rho relaxation times of fat and water. To compare bone with a sample matrix with magnetic susceptibility interfaces, phantoms consisting of glass beads with different diameters in oil or water were used. The behavior of T-2 measured with different sequences and T-1 rho at different magnitudes of spin-lock fields were characterized, and relaxation times were correlated with structural parameters. T-2 and T-1. showed significant associations with structural bone parameters. Strongest linear correlations (r = 0.81, p < 0.01) were established between R-1 rho (1/T-1 rho) of fat component and structural model index. For glass beads, the behavior of T-2 and T-1 rho was similar to that of the water compartment of bone marrow. The present results suggest feasibility of spectroscopic NMR measurements to assess trabecular structure. However, further studies are required to determine the sensitivity of this approach to fat content of bone marrow and to lower the field strengths used in clinical devices.
机译:已经开发了磁共振成像(MRI)技术,用于无创评估小梁骨的结构特性。然而,与小梁尺寸相比,这些测量具有相对较长的采集时间和较低的分辨率。弛豫时间的光谱测量可以用于更详细,更快地评估骨髓的弛豫特性,还可以提供有关骨小梁结构的替代信息。在本研究中,使用光谱NMR(核磁共振)方法对牛小梁骨进行了研究,以确定用micro-CT和T-2,Carr-Purcell T-2和T-1 rho弛豫时间测量的结构参数之间的关系。脂肪和水。为了将骨骼与具有磁化率界面的样品基质进行比较,使用了由在油或水中具有不同直径的玻璃珠组成的体模。表征了不同序列的T-2和在不同自旋锁磁场强度下的T-1 rho行为,弛豫时间与结构参数相关。 T-2和T-1。表现出与骨结构参数的显着关联。在脂肪成分的R-1 rho(1 / T-1 rho)与结构模型指数之间建立了最强的线性相关性(r = 0.81,p <0.01)。对于玻璃珠,T-2和T-1 rho的行为与骨髓水室的行为相似。目前的结果表明光谱核磁共振测量评估骨小梁结构的可行性。但是,需要进一步的研究来确定这种方法对骨髓脂肪含量的敏感性,并降低临床设备中使用的场强。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号