首页> 外文期刊>Physics in medicine and biology. >Testing the quality of images for permanent magnet desktop MRI systems using specially designed phantoms
【24h】

Testing the quality of images for permanent magnet desktop MRI systems using specially designed phantoms

机译:使用特殊设计的体模测试永磁台式MRI系统的图像质量

获取原文
获取原文并翻译 | 示例

摘要

Our aim was to measure the performance of desktop magnetic resonance imaging (MRI) systems using specially designed phantoms, by testing imaging parameters and analysing the imaging quality. We designed multifunction phantoms with diameters of 18 and 60 mm for desktop MRI scanners in accordance with the American Association of Physicists in Medicine (AAPM) report no. 28. We scanned the phantoms with three permanent magnet 0.5 T desktop MRI systems, measured the MRI image parameters, and analysed imaging quality by comparing the data with the AAPM criteria and Chinese national standards. Image parameters included: resonance frequency, high contrast spatial resolution, low contrast object detectability, slice thickness, geometrical distortion, signal-to-noise ratio (SNR), and image uniformity. The image parameters of three desktop MRI machines could be measured using our specially designed phantoms, and most parameters were in line with MRI quality control criterion, including: resonance frequency, high contrast spatial resolution, low contrast object detectability, slice thickness, geometrical distortion, image uniformity and slice position accuracy. However, SNR was significantly lower than in some references. The imaging test and quality control are necessary for desktop MRI systems, and should be performed with the applicable phantom and corresponding standards.
机译:我们的目标是通过测试成像参数并分析成像质量,使用专门设计的体模来测量台式磁共振成像(MRI)系统的性能。根据美国医学物理学家协会(AAPM)报告第1号,我们为台式MRI扫描仪设计了直径分别为18和60 mm的多功能体模。 28.我们使用三个永磁体0.5 T台式MRI系统扫描了人体模型,测量了MRI图像参数,并通过将数据与AAPM标准和中国国家标准相比较来分析了成像质量。图像参数包括:共振频率,高对比度空间分辨率,低对比度对象可检测性,切片厚度,几何失真,信噪比(SNR)和图像均匀性。可以使用我们专门设计的体模来测量三台台式MRI机器的图像参数,并且大多数参数都符合MRI质量控制标准,包括:共振频率,高对比度空间分辨率,低对比度物体可检测性,切片厚度,几何变形,图像均匀性和切片位置精度。但是,SNR明显低于某些参考文献。台式MRI系统必须进行成像测试和质量控制,并应使用适用的体模和相应的标准进行。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号